1.Endoscopy Center, Zhongshan Hospital, Fudan University, Shanghai 200032, China;2.Department of Infection Management, Zhongshan Hospital, Fudan University, Shanghai 200032, China;3.Department of Infection Ma-nagement, Minhang Hospital, Fudan University, Shanghai 201199, China;4.Department of Infectious Diseases, Zhongshan Hospital, Fudan University, Shanghai 200032, China

Clc Number:

+.2]]>

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Comments
    Abstract:

    Objective To discuss the disinfection effect of persistent hypochlorous acid supply to final rinse water for endoscope. Methods Water specimens of final rinse water for endoscope were collected at baseline, and identified by microbial culture and metagenomic next-generation sequencing (mNGS). After introducing a slightly acidic hypochlorous acid water generator, water specimens were collected at a specific time after installation for culture and mNGS analysis. Results A total of 27 water specimens were collected before installation, and 10 (37.03%) were qualified. Aquatic bacteria such as Rhizobiumradiobacter, Stenotrophomonas maltophilia, Microbacterium flavum, Cupriavidus, Methylobacterium fujisawaense and Sphingomonas paucimobilis can be detected in unqualified water specimens, 54 water specimens were collected after installation, and all of them were qualified. mNGS shows that abundant sequences of Ralstonia Pichia, Cupriavidus metallidurans, Moraxella osloensis, Rhizobium japonicum, Aquatic methylbacterium, Mycobacterium Gordon and Sphingosine neobacterium could be detected in water specimens before installation, but only background bacteria in mNGS were found after disinfection. Conclusion Persistent hypochlorous acid supply can effectively remove bacteria and reduce biofilm formation in water pipeline, improving the quality of final rinse water for endoscope.

    Reference
    Related
Get Citation

王萍,韩梦鸽,沈国锋,等.持续性次氯酸消毒对内镜终末漂洗水的消毒效果[J].中国感染控制杂志英文版,2022,(7):617-622. DOI:10.12138/j. issn.1671-9638.20222955.
Ping WANG, Meng-ge HAN, Guo-feng SHEN, et al. Disinfection effect of persistent hypochlorous acid on final rinsing water for endoscope[J]. Chin J Infect Control, 2022,(7):617-622. DOI:10.12138/j. issn.1671-9638.20222955.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 30,2022
  • Revised:
  • Adopted:
  • Online: August 01,2022
  • Published: