Abstract:Objective To explore the effect of combined application of the PDCA cycle and root cause analysis (RCA) on improving the cleaning quality of da Vinci robotic surgical arms.Methods The study was conducted at the central sterile supply department of a grade A tertiary hospital in Hunan Province. All batches of da Vinci robotic surgical arms cleaning within pre-defined time windows were consecutively included. A total of 112 instrument-cleaning instances from May 1 to 31, 2024, were included as the preintervention baseline; 123 instances from June 17 to July 16, 2024, were used for verification of the actual causes; and 232 instances from August 1 to September 30, 2024, were included for outcome evaluation. The PDCA cycle served as the overall improvement framework, with RCA embedded in the planning phase. Brainstorming, fishbone diagrams, onsite observation, and video replay were employed to identify and verify the actual causes. Improvements were implemented across three dimensions: procedures, equipment, and personnel.Results The cleaning pass rate increased from 82.14% (92/112) before improvement to 96.98% (225/232) after improvement, with a statistically significant difference (χ2=22.999,P<0.001). The monthly number of unqualified items due to nonstandardized procedures, lack of dedicated cleaning equipment and programs, and insufficient work experience decreased from 9, 9, and 6, to 2, 0, and 1, respectively. The overall mean score for self-assessed abilities among the 10 members of the dedicated project team increased from (2.85±0.14) to (4.61±0.17) points.Conclusion The combined application of the PDCA cycle and RCA can link rootcause identification with closedloop improvement and effectively improve the cleaning quality of da Vinci robotic surgical arms. It is recommended to continuously consolidate the improvements through procedure standardization, provision of dedicated equipment, and tiered training.