Investigation and intervention effectiveness in a standard-exceeding event of microbial monitoring of pediatric ultrathin flexible bronchoscopes
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 文章评论
    摘要:

    目的 对某三级甲等妇幼保健院一起儿科超细软式支气管镜(以下简称软镜)微生物监测超标事件开展调查与干预,为提高软镜清洗与消毒质量、减少因软镜污染而导致的医院感染提供依据。方法 联系厂家召回微生物超标软镜,使用管道镜探查软镜管道内部,并进行测漏;现场对终末漂洗用水系统采样;通过现场调研了解工作人员软镜处置流程及使用情况;制定干预措施。结果 主要问题有:①软镜管腔表面有明显划伤及表皮翘起,测漏结果显示微漏;②终末漂洗用水微生物检测超标;③部分软镜处置人员上岗前未经规范培训与考核,软镜干燥时间不足,刷洗软镜动作粗暴。针对存在问题,厂家更换软镜外皮及漂洗用水发生器过滤装置,并在终末漂洗用水出水口加装过滤器,医院采取加强对软镜处置人员的培训、考核与监管等措施后,再次检测结果合格。结论 本次软镜微生物超标可能由软镜表皮损伤导致生物膜形成、终末漂洗用水污染、人员操作不规范等原因所致。为预防软镜相关医院感染发生,医疗机构应定期检查使用中的软式软镜管腔并测漏,对损伤较为严重的软镜及时维修。此外,还需关注漂洗用水质量、供水管路消毒、处置人员培训与考核等情况。

    Abstract:

    Objective To investigate and intervene in a standard-exceeding event of microbial monitoring of pediatric ultrathin flexible bronchoscopes in a tertiary first-class maternity and child health care hospital, and to provide a basis for improving the cleaning and disinfection quality of flexible bronchoscopes and reducing healthcare-associated infection (HAI) caused by contamination of flexible bronchoscopes. Methods Flexible bronchoscopes with microbial contamination were recalled through contacting the manufacturer. The internal channels of the flexible bronchoscopes were examined using borescope, and leakage was tested. On-site sampling of the final rinse water system was conducted. The usage and disposal procedures of flexible bronchoscopes implemented by staff were investigated through on-site survey. Intervention measures were developed. Results The main problems were: 1 There were visible scratches and epidermal lifting on the tube surface of flexible bronchoscopes, with leakage testing indicating minor leakage. 2 Microbial detection of the final rinse water exceeded standards. 3 Some flexible bronchoscopes handling personnel lacked standardized pre-service training and assessment, leading to inadequate drying time and rough brushing of the bronchoscopes. In response to the existing problems, the manufacturer replaced the flexible bronchoscopes outer sheath and the filter for the rinse water generator, and installed a filter at the outlet of the final rinse water. After the hospital strengthened training, assessment, and supervision of the personnel handling flexible bronchoscopes, the re-inspection results were qualified. Conclusion The microbial count exceeding the standard for flexible bronchoscopes may be attributed to biofilm formation resulting from surface damage, contamination of final rinse water, and improper personnel handling. To prevent HAI related to flexible bronchoscopes, medical institutions should regularly inspect the lumens of in-use flexible bronchoscopes and test leakage, promptly repair those with significant damage. In addition, it is necessary to pay attention to the quality of rinse water, disinfection of water supply lines, as well as training and assessment of disposal personnel.

    参考文献
    相似文献
引用本文

张翔,付路,郭艳,等.儿科超细软式支气管镜微生物监测超标事件调查与干预效果研究[J]. 中国感染控制杂志,2026,25(7):1042-1048. DOI:10.12138/j. issn.1671-9638.20263070.
ZHANG Xiang, FU Lu, GUO Yan, et al. Investigation and intervention effectiveness in a standard-exceeding event of microbial monitoring of pediatric ultrathin flexible bronchoscopes[J]. Chin J Infect Control, 2026,25(7):1042-1048. DOI:10.12138/j. issn.1671-9638.20263070.

复制
分享
文章指标
  • 摘要阅读次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2025-10-30
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2026-07-27
  • 出版日期: 2026-07-28